11 After a `quick repair of the PSU of my Tek 2445 |
11 After a `quick repair of the PSU of my Tek 2445 |
12 <{static}/TeK2445.rst>`_, I've started a more systematic replacement |
12 <{filename}/TeK2445.rst>`_, I've started a more systematic replacement |
13 of the capacitors. |
13 of the capacitors. |
14 |
14 |
15 .. image:: {static}/images/tek2445/psu_caps_replaced.jpg |
15 .. image:: {static}/images/tek2445/psu_caps_replaced.jpg |
16 :alt: Tek2445 PSU Caps being replaced. |
16 :alt: Tek2445 PSU Caps being replaced. |
17 |
17 |
57 exact specifications of the device; the service manual do not give the |
57 exact specifications of the device; the service manual do not give the |
58 value of the resistor in parallel with the 2.5µH inductance. |
58 value of the resistor in parallel with the 2.5µH inductance. |
59 |
59 |
60 .. image:: {static}/images/tek2445/dead_transistors.jpg |
60 .. image:: {static}/images/tek2445/dead_transistors.jpg |
61 :alt: Q1050, 1060 and Q1070 are dead. |
61 :alt: Q1050, 1060 and Q1070 are dead. |
62 |
62 |
63 For now, I've picked up a 3W 100Ω resistor and I've made a self by |
63 For now, I've picked up a 3W 100Ω resistor and I've made a self by |
64 hand. As I don't have a RLC meter, I don't known its exact value so |
64 hand. As I don't have a RLC meter, I don't known its exact value so |
65 I've took a few measures of its impedance at several frequencies, and |
65 I've took a few measures of its impedance at several frequencies, and |
66 I estimate the inductance to be around 4 µH. I have no idea whether |
66 I estimate the inductance to be around 4 µH. I have no idea whether |
67 this higher value might be a problem... But I've no idea either if the |
67 this higher value might be a problem... But I've no idea either if the |